Thermistor Standards
Probes Models 5640–5644
|
Features
|
|
Accuracy to ±0.001°C
|
|
Affordable system accuracy to ±0.004°C or better
|
|
NIST-traceable calibration included
|
|
Specifications
|
|
Model
|
Dia x Length
|
Range
|
Stability
°C/Year
|
Accuracy
0–60°C
60–100°C
|
Wires
|
|
5640
|
.25" x 9"
(6.35 x 229 mm)
|
0°C–60°C
|
±0.005°C
|
±0.0015°C
|
|
4
|
|
5641
|
.125" x 4.5"
(3.18 x 114 mm)
|
0°C–60°C
|
±0.002°C
|
±0.001°C
|
|
4
|
|
5642
|
.125" x 9"
(3.18 x 229 mm)
|
0°C–60°C
|
±0.002°C
|
±0.001°C
|
|
4
|
|
5643
|
.125" x 4.5"
(3.18 x 114 mm)
|
0°C–100°C
|
±0.005°C
|
±0.0015°C
|
±0.0025°C
|
4
|
|
5644
|
.125" x 9"
(3.18 x 229 mm)
|
0°C–100°C
|
±0.005°C
|
±0.0015°C
|
±0.0025°C
|
4
|
If you want a high-accuracy
probe with excellent stability at a great price, the Model 5640-series
Thermistor Standards Probes give you all three in a great package. Why
pay for an SPRT when you can get ±0.001°C accuracy from 0°C to 60°C in
a calibrated thermistor probe for about one-third the cost of an uncalibrated
SPRT alone?
Each probe uses
an ultra-stable glass thermistor enclosed in a thin-wall stainless steel
tube. The basic semiconductor element is a bead of manganese, nickel,
and cobalt oxides mounted on 0.1 mm platinum wires. For long-term stability,
the thermistor is aged at various temperatures for 16 weeks. During the
aging process, verification of the probe’s stability is done to ensure
performance to published specs.
The 5640, 5641,
and 5642 thermistor probes are designed for the temperature range of 0°C
to 60°C. The 5643 and 5644 probes span the 0°C to 100°C temperature range.
They offer stability of either ±0.002°C or ±0.005°C. These stability levels
are guaranteed for one full year.
Precision calibration,
traceable to NIST, is provided with each probe. A computer-generated table
in increments of 0.01°C is furnished with each calibration based on the
formula:
R=exp(A + B/T + C/T2 + D/T3)
The constants for
the formula are obtained from a polynomial regression performed on the
calibration data obtained. Over the range of 0°C to 60°C, calibration
is performed at the triple point of water (0.01°C) and 15°C, 25°C, 30°C,
37°C, 50°C and 60°C. For the 0°C to 100°C temperature range, the additional
calibration points of 80°C and 100°C are used.
Each probe is individually
calibrated and includes a report of calibration from the manufacturer.
Contact Hart for calibration in Hart’s NVLAP accredited lab.
Thermistor standards
are rugged, precision sensors suitable for use as secondary or working
temperature standards for laboratory metrology applications. Because they
generally are not affected by shock and vibration, you can use them in
the most difficult field environments without worrying about calibration
integrity.
Combine these probes
with Hart’s Model 1560 Black Stack thermometer to read directly
in °C, °F, or K. This combination gives you resolution of 0.0001 degrees
and total system accuracy is better than ±0.004°C. Combined with Hart’s
Model 1575 Super-Thermometer, the 5640 gives you an overall system accuracy
of ±0.002°C.
Compare the cost
of a 5640 calibrated probe and a Black Stack thermometer to the
cost of one uncalibrated SPRT. Between 0°C and 100°C, nothing beats the
value of the Model 5640 Thermistor Standards Probe.
|